1
0
Weather_ESP/main/bme280.c
Christian Loch b12deb401c Imported Bosch BME280 driver
Added ESP-IDF example for I2C bus
2020-02-04 00:00:54 +01:00

1497 lines
48 KiB
C

/**
* Copyright (c) 2020 Bosch Sensortec GmbH. All rights reserved.
*
* BSD-3-Clause
*
* Redistribution and use in source and binary forms, with or without
* modification, are permitted provided that the following conditions are met:
*
* 1. Redistributions of source code must retain the above copyright
* notice, this list of conditions and the following disclaimer.
*
* 2. Redistributions in binary form must reproduce the above copyright
* notice, this list of conditions and the following disclaimer in the
* documentation and/or other materials provided with the distribution.
*
* 3. Neither the name of the copyright holder nor the names of its
* contributors may be used to endorse or promote products derived from
* this software without specific prior written permission.
*
* THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS
* "AS IS" AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT
* LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
* FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE
* COPYRIGHT HOLDER OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT,
* INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES
* (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR
* SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION)
* HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT,
* STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING
* IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE
* POSSIBILITY OF SUCH DAMAGE.
*
* @file bme280.c
* @date 21/01/2020
* @version 3.4.2
*
*/
/*! @file bme280.c
* @brief Sensor driver for BME280 sensor
*/
#include "bme280.h"
/**\name Internal macros */
/* To identify osr settings selected by user */
#define OVERSAMPLING_SETTINGS UINT8_C(0x07)
/* To identify filter and standby settings selected by user */
#define FILTER_STANDBY_SETTINGS UINT8_C(0x18)
/*!
* @brief This internal API puts the device to sleep mode.
*
* @param[in] dev : Structure instance of bme280_dev.
*
* @return Result of API execution status.
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t put_device_to_sleep(const struct bme280_dev *dev);
/*!
* @brief This internal API writes the power mode in the sensor.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[in] sensor_mode : Variable which contains the power mode to be set.
*
* @return Result of API execution status.
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t write_power_mode(uint8_t sensor_mode, const struct bme280_dev *dev);
/*!
* @brief This internal API is used to validate the device pointer for
* null conditions.
*
* @param[in] dev : Structure instance of bme280_dev.
*
* @return Result of API execution status
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t null_ptr_check(const struct bme280_dev *dev);
/*!
* @brief This internal API interleaves the register address between the
* register data buffer for burst write operation.
*
* @param[in] reg_addr : Contains the register address array.
* @param[out] temp_buff : Contains the temporary buffer to store the
* register data and register address.
* @param[in] reg_data : Contains the register data to be written in the
* temporary buffer.
* @param[in] len : No of bytes of data to be written for burst write.
*/
static void interleave_reg_addr(const uint8_t *reg_addr, uint8_t *temp_buff, const uint8_t *reg_data, uint8_t len);
/*!
* @brief This internal API reads the calibration data from the sensor, parse
* it and store in the device structure.
*
* @param[in] dev : Structure instance of bme280_dev.
*
* @return Result of API execution status
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t get_calib_data(struct bme280_dev *dev);
/*!
* @brief This internal API is used to parse the temperature and
* pressure calibration data and store it in the device structure.
*
* @param[out] dev : Structure instance of bme280_dev to store the calib data.
* @param[in] reg_data : Contains the calibration data to be parsed.
*/
static void parse_temp_press_calib_data(const uint8_t *reg_data, struct bme280_dev *dev);
/*!
* @brief This internal API is used to parse the humidity calibration data
* and store it in device structure.
*
* @param[out] dev : Structure instance of bme280_dev to store the calib data.
* @param[in] reg_data : Contains calibration data to be parsed.
*/
static void parse_humidity_calib_data(const uint8_t *reg_data, struct bme280_dev *dev);
#ifdef BME280_FLOAT_ENABLE
/*!
* @brief This internal API is used to compensate the raw pressure data and
* return the compensated pressure data in double data type.
*
* @param[in] uncomp_data : Contains the uncompensated pressure data.
* @param[in] calib_data : Pointer to the calibration data structure.
*
* @return Compensated pressure data.
* @retval Compensated pressure data in double.
*/
static double compensate_pressure(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data);
/*!
* @brief This internal API is used to compensate the raw humidity data and
* return the compensated humidity data in double data type.
*
* @param[in] uncomp_data : Contains the uncompensated humidity data.
* @param[in] calib_data : Pointer to the calibration data structure.
*
* @return Compensated humidity data.
* @retval Compensated humidity data in double.
*/
static double compensate_humidity(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data);
/*!
* @brief This internal API is used to compensate the raw temperature data and
* return the compensated temperature data in double data type.
*
* @param[in] uncomp_data : Contains the uncompensated temperature data.
* @param[in] calib_data : Pointer to calibration data structure.
*
* @return Compensated temperature data.
* @retval Compensated temperature data in double.
*/
static double compensate_temperature(const struct bme280_uncomp_data *uncomp_data,
struct bme280_calib_data *calib_data);
#else
/*!
* @brief This internal API is used to compensate the raw temperature data and
* return the compensated temperature data in integer data type.
*
* @param[in] uncomp_data : Contains the uncompensated temperature data.
* @param[in] calib_data : Pointer to calibration data structure.
*
* @return Compensated temperature data.
* @retval Compensated temperature data in integer.
*/
static int32_t compensate_temperature(const struct bme280_uncomp_data *uncomp_data,
struct bme280_calib_data *calib_data);
/*!
* @brief This internal API is used to compensate the raw pressure data and
* return the compensated pressure data in integer data type.
*
* @param[in] uncomp_data : Contains the uncompensated pressure data.
* @param[in] calib_data : Pointer to the calibration data structure.
*
* @return Compensated pressure data.
* @retval Compensated pressure data in integer.
*/
static uint32_t compensate_pressure(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data);
/*!
* @brief This internal API is used to compensate the raw humidity data and
* return the compensated humidity data in integer data type.
*
* @param[in] uncomp_data : Contains the uncompensated humidity data.
* @param[in] calib_data : Pointer to the calibration data structure.
*
* @return Compensated humidity data.
* @retval Compensated humidity data in integer.
*/
static uint32_t compensate_humidity(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data);
#endif
/*!
* @brief This internal API is used to identify the settings which the user
* wants to modify in the sensor.
*
* @param[in] sub_settings : Contains the settings subset to identify particular
* group of settings which the user is interested to change.
* @param[in] desired_settings : Contains the user specified settings.
*
* @return Indicates whether user is interested to modify the settings which
* are related to sub_settings.
* @retval True -> User wants to modify this group of settings
* @retval False -> User does not want to modify this group of settings
*/
static uint8_t are_settings_changed(uint8_t sub_settings, uint8_t desired_settings);
/*!
* @brief This API sets the humidity oversampling settings of the sensor.
*
* @param[in] dev : Structure instance of bme280_dev.
*
* @return Result of API execution status
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t set_osr_humidity_settings(const struct bme280_settings *settings, const struct bme280_dev *dev);
/*!
* @brief This internal API sets the oversampling settings for pressure,
* temperature and humidity in the sensor.
*
* @param[in] desired_settings : Variable used to select the settings which
* are to be set.
* @param[in] dev : Structure instance of bme280_dev.
*
* @return Result of API execution status
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t set_osr_settings(uint8_t desired_settings,
const struct bme280_settings *settings,
const struct bme280_dev *dev);
/*!
* @brief This API sets the pressure and/or temperature oversampling settings
* in the sensor according to the settings selected by the user.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[in] desired_settings: variable to select the pressure and/or
* temperature oversampling settings.
*
* @return Result of API execution status
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t set_osr_press_temp_settings(uint8_t desired_settings,
const struct bme280_settings *settings,
const struct bme280_dev *dev);
/*!
* @brief This internal API fills the pressure oversampling settings provided by
* the user in the data buffer so as to write in the sensor.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[out] reg_data : Variable which is filled according to the pressure
* oversampling data provided by the user.
*/
static void fill_osr_press_settings(uint8_t *reg_data, const struct bme280_settings *settings);
/*!
* @brief This internal API fills the temperature oversampling settings provided
* by the user in the data buffer so as to write in the sensor.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[out] reg_data : Variable which is filled according to the temperature
* oversampling data provided by the user.
*/
static void fill_osr_temp_settings(uint8_t *reg_data, const struct bme280_settings *settings);
/*!
* @brief This internal API sets the filter and/or standby duration settings
* in the sensor according to the settings selected by the user.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[in] desired_settings : variable to select the filter and/or
* standby duration settings.
*
* @return Result of API execution status
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t set_filter_standby_settings(uint8_t desired_settings,
const struct bme280_settings *settings,
const struct bme280_dev *dev);
/*!
* @brief This internal API fills the filter settings provided by the user
* in the data buffer so as to write in the sensor.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[out] reg_data : Variable which is filled according to the filter
* settings data provided by the user.
*/
static void fill_filter_settings(uint8_t *reg_data, const struct bme280_settings *settings);
/*!
* @brief This internal API fills the standby duration settings provided by the
* user in the data buffer so as to write in the sensor.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[out] reg_data : Variable which is filled according to the standby
* settings data provided by the user.
*/
static void fill_standby_settings(uint8_t *reg_data, const struct bme280_settings *settings);
/*!
* @brief This internal API parse the oversampling(pressure, temperature
* and humidity), filter and standby duration settings and store in the
* device structure.
*
* @param[out] dev : Structure instance of bme280_dev.
* @param[in] reg_data : Register data to be parsed.
*/
static void parse_device_settings(const uint8_t *reg_data, struct bme280_settings *settings);
/*!
* @brief This internal API reloads the already existing device settings in the
* sensor after soft reset.
*
* @param[in] dev : Structure instance of bme280_dev.
* @param[in] settings : Pointer variable which contains the settings to
* be set in the sensor.
*
* @return Result of API execution status
* @retval zero -> Success / +ve value -> Warning / -ve value -> Error
*/
static int8_t reload_device_settings(const struct bme280_settings *settings, const struct bme280_dev *dev);
/****************** Global Function Definitions *******************************/
/*!
* @brief This API is the entry point.
* It reads the chip-id and calibration data from the sensor.
*/
int8_t bme280_init(struct bme280_dev *dev)
{
int8_t rslt;
/* chip id read try count */
uint8_t try_count = 5;
uint8_t chip_id = 0;
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
/* Proceed if null check is fine */
if (rslt == BME280_OK)
{
while (try_count)
{
/* Read the chip-id of bme280 sensor */
rslt = bme280_get_regs(BME280_CHIP_ID_ADDR, &chip_id, 1, dev);
/* Check for chip id validity */
if ((rslt == BME280_OK) && (chip_id == BME280_CHIP_ID))
{
dev->chip_id = chip_id;
/* Reset the sensor */
rslt = bme280_soft_reset(dev);
if (rslt == BME280_OK)
{
/* Read the calibration data */
rslt = get_calib_data(dev);
}
break;
}
/* Wait for 1 ms */
dev->delay_ms(1);
--try_count;
}
/* Chip id check failed */
if (!try_count)
{
rslt = BME280_E_DEV_NOT_FOUND;
}
}
return rslt;
}
/*!
* @brief This API reads the data from the given register address of the sensor.
*/
int8_t bme280_get_regs(uint8_t reg_addr, uint8_t *reg_data, uint16_t len, const struct bme280_dev *dev)
{
int8_t rslt;
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
/* Proceed if null check is fine */
if (rslt == BME280_OK)
{
/* If interface selected is SPI */
if (dev->intf != BME280_I2C_INTF)
{
reg_addr = reg_addr | 0x80;
}
/* Read the data */
rslt = dev->read(dev->dev_id, reg_addr, reg_data, len);
/* Check for communication error */
if (rslt != BME280_OK)
{
rslt = BME280_E_COMM_FAIL;
}
}
return rslt;
}
/*!
* @brief This API writes the given data to the register address
* of the sensor.
*/
int8_t bme280_set_regs(uint8_t *reg_addr, const uint8_t *reg_data, uint8_t len, const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t temp_buff[20]; /* Typically not to write more than 10 registers */
if (len > 10)
{
len = 10;
}
uint16_t temp_len;
uint8_t reg_addr_cnt;
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
/* Check for arguments validity */
if ((rslt == BME280_OK) && (reg_addr != NULL) && (reg_data != NULL))
{
if (len != 0)
{
temp_buff[0] = reg_data[0];
/* If interface selected is SPI */
if (dev->intf != BME280_I2C_INTF)
{
for (reg_addr_cnt = 0; reg_addr_cnt < len; reg_addr_cnt++)
{
reg_addr[reg_addr_cnt] = reg_addr[reg_addr_cnt] & 0x7F;
}
}
/* Burst write mode */
if (len > 1)
{
/* Interleave register address w.r.t data for
* burst write
*/
interleave_reg_addr(reg_addr, temp_buff, reg_data, len);
temp_len = ((len * 2) - 1);
}
else
{
temp_len = len;
}
rslt = dev->write(dev->dev_id, reg_addr[0], temp_buff, temp_len);
/* Check for communication error */
if (rslt != BME280_OK)
{
rslt = BME280_E_COMM_FAIL;
}
}
else
{
rslt = BME280_E_INVALID_LEN;
}
}
else
{
rslt = BME280_E_NULL_PTR;
}
return rslt;
}
/*!
* @brief This API sets the oversampling, filter and standby duration
* (normal mode) settings in the sensor.
*/
int8_t bme280_set_sensor_settings(uint8_t desired_settings, const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t sensor_mode;
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
/* Proceed if null check is fine */
if (rslt == BME280_OK)
{
rslt = bme280_get_sensor_mode(&sensor_mode, dev);
if ((rslt == BME280_OK) && (sensor_mode != BME280_SLEEP_MODE))
{
rslt = put_device_to_sleep(dev);
}
if (rslt == BME280_OK)
{
/* Check if user wants to change oversampling
* settings
*/
if (are_settings_changed(OVERSAMPLING_SETTINGS, desired_settings))
{
rslt = set_osr_settings(desired_settings, &dev->settings, dev);
}
/* Check if user wants to change filter and/or
* standby settings
*/
if ((rslt == BME280_OK) && are_settings_changed(FILTER_STANDBY_SETTINGS, desired_settings))
{
rslt = set_filter_standby_settings(desired_settings, &dev->settings, dev);
}
}
}
return rslt;
}
/*!
* @brief This API gets the oversampling, filter and standby duration
* (normal mode) settings from the sensor.
*/
int8_t bme280_get_sensor_settings(struct bme280_dev *dev)
{
int8_t rslt;
uint8_t reg_data[4];
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
/* Proceed if null check is fine */
if (rslt == BME280_OK)
{
rslt = bme280_get_regs(BME280_CTRL_HUM_ADDR, reg_data, 4, dev);
if (rslt == BME280_OK)
{
parse_device_settings(reg_data, &dev->settings);
}
}
return rslt;
}
/*!
* @brief This API sets the power mode of the sensor.
*/
int8_t bme280_set_sensor_mode(uint8_t sensor_mode, const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t last_set_mode;
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
if (rslt == BME280_OK)
{
rslt = bme280_get_sensor_mode(&last_set_mode, dev);
/* If the sensor is not in sleep mode put the device to sleep
* mode
*/
if ((rslt == BME280_OK) && (last_set_mode != BME280_SLEEP_MODE))
{
rslt = put_device_to_sleep(dev);
}
/* Set the power mode */
if (rslt == BME280_OK)
{
rslt = write_power_mode(sensor_mode, dev);
}
}
return rslt;
}
/*!
* @brief This API gets the power mode of the sensor.
*/
int8_t bme280_get_sensor_mode(uint8_t *sensor_mode, const struct bme280_dev *dev)
{
int8_t rslt;
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
if (rslt == BME280_OK)
{
/* Read the power mode register */
rslt = bme280_get_regs(BME280_PWR_CTRL_ADDR, sensor_mode, 1, dev);
/* Assign the power mode in the device structure */
*sensor_mode = BME280_GET_BITS_POS_0(*sensor_mode, BME280_SENSOR_MODE);
}
return rslt;
}
/*!
* @brief This API performs the soft reset of the sensor.
*/
int8_t bme280_soft_reset(const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t reg_addr = BME280_RESET_ADDR;
uint8_t status_reg = 0;
uint8_t try_run = 5;
/* 0xB6 is the soft reset command */
uint8_t soft_rst_cmd = BME280_SOFT_RESET_COMMAND;
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
/* Proceed if null check is fine */
if (rslt == BME280_OK)
{
/* Write the soft reset command in the sensor */
rslt = bme280_set_regs(&reg_addr, &soft_rst_cmd, 1, dev);
if (rslt == BME280_OK)
{
/* If NVM not copied yet, Wait for NVM to copy */
do
{
/* As per data sheet - Table 1, startup time is 2 ms. */
dev->delay_ms(2);
rslt = bme280_get_regs(BME280_STATUS_REG_ADDR, &status_reg, 1, dev);
} while ((rslt == BME280_OK) && (try_run--) && (status_reg & BME280_STATUS_IM_UPDATE));
if (status_reg & BME280_STATUS_IM_UPDATE)
{
rslt = BME280_E_NVM_COPY_FAILED;
}
}
}
return rslt;
}
/*!
* @brief This API reads the pressure, temperature and humidity data from the
* sensor, compensates the data and store it in the bme280_data structure
* instance passed by the user.
*/
int8_t bme280_get_sensor_data(uint8_t sensor_comp, struct bme280_data *comp_data, struct bme280_dev *dev)
{
int8_t rslt;
/* Array to store the pressure, temperature and humidity data read from
* the sensor
*/
uint8_t reg_data[BME280_P_T_H_DATA_LEN] = { 0 };
struct bme280_uncomp_data uncomp_data = { 0 };
/* Check for null pointer in the device structure*/
rslt = null_ptr_check(dev);
if ((rslt == BME280_OK) && (comp_data != NULL))
{
/* Read the pressure and temperature data from the sensor */
rslt = bme280_get_regs(BME280_DATA_ADDR, reg_data, BME280_P_T_H_DATA_LEN, dev);
if (rslt == BME280_OK)
{
/* Parse the read data from the sensor */
bme280_parse_sensor_data(reg_data, &uncomp_data);
/* Compensate the pressure and/or temperature and/or
* humidity data from the sensor
*/
rslt = bme280_compensate_data(sensor_comp, &uncomp_data, comp_data, &dev->calib_data);
}
}
else
{
rslt = BME280_E_NULL_PTR;
}
return rslt;
}
/*!
* @brief This API is used to parse the pressure, temperature and
* humidity data and store it in the bme280_uncomp_data structure instance.
*/
void bme280_parse_sensor_data(const uint8_t *reg_data, struct bme280_uncomp_data *uncomp_data)
{
/* Variables to store the sensor data */
uint32_t data_xlsb;
uint32_t data_lsb;
uint32_t data_msb;
/* Store the parsed register values for pressure data */
data_msb = (uint32_t)reg_data[0] << 12;
data_lsb = (uint32_t)reg_data[1] << 4;
data_xlsb = (uint32_t)reg_data[2] >> 4;
uncomp_data->pressure = data_msb | data_lsb | data_xlsb;
/* Store the parsed register values for temperature data */
data_msb = (uint32_t)reg_data[3] << 12;
data_lsb = (uint32_t)reg_data[4] << 4;
data_xlsb = (uint32_t)reg_data[5] >> 4;
uncomp_data->temperature = data_msb | data_lsb | data_xlsb;
/* Store the parsed register values for humidity data */
data_msb = (uint32_t)reg_data[6] << 8;
data_lsb = (uint32_t)reg_data[7];
uncomp_data->humidity = data_msb | data_lsb;
}
/*!
* @brief This API is used to compensate the pressure and/or
* temperature and/or humidity data according to the component selected
* by the user.
*/
int8_t bme280_compensate_data(uint8_t sensor_comp,
const struct bme280_uncomp_data *uncomp_data,
struct bme280_data *comp_data,
struct bme280_calib_data *calib_data)
{
int8_t rslt = BME280_OK;
if ((uncomp_data != NULL) && (comp_data != NULL) && (calib_data != NULL))
{
/* Initialize to zero */
comp_data->temperature = 0;
comp_data->pressure = 0;
comp_data->humidity = 0;
/* If pressure or temperature component is selected */
if (sensor_comp & (BME280_PRESS | BME280_TEMP | BME280_HUM))
{
/* Compensate the temperature data */
comp_data->temperature = compensate_temperature(uncomp_data, calib_data);
}
if (sensor_comp & BME280_PRESS)
{
/* Compensate the pressure data */
comp_data->pressure = compensate_pressure(uncomp_data, calib_data);
}
if (sensor_comp & BME280_HUM)
{
/* Compensate the humidity data */
comp_data->humidity = compensate_humidity(uncomp_data, calib_data);
}
}
else
{
rslt = BME280_E_NULL_PTR;
}
return rslt;
}
/*!
* @brief This API is used to calculate the maximum delay in milliseconds required for the
* temperature/pressure/humidity(which ever at enabled) measurement to complete.
*/
uint32_t bme280_cal_meas_delay(const struct bme280_settings *settings)
{
uint32_t max_delay;
uint8_t temp_osr;
uint8_t pres_osr;
uint8_t hum_osr;
/*Array to map OSR config register value to actual OSR */
uint8_t osr_sett_to_act_osr[] = { 0, 1, 2, 4, 8, 16 };
/* Mapping osr settings to the actual osr values e.g. 0b101 -> osr X16 */
if (settings->osr_t <= 5)
{
temp_osr = osr_sett_to_act_osr[settings->osr_t];
}
else
{
temp_osr = 16;
}
if (settings->osr_p <= 5)
{
pres_osr = osr_sett_to_act_osr[settings->osr_p];
}
else
{
pres_osr = 16;
}
if (settings->osr_h <= 5)
{
hum_osr = osr_sett_to_act_osr[settings->osr_h];
}
else
{
hum_osr = 16;
}
max_delay =
(uint32_t)((BME280_MEAS_OFFSET + (BME280_MEAS_DUR * temp_osr) +
((BME280_MEAS_DUR * pres_osr) + BME280_PRES_HUM_MEAS_OFFSET) +
((BME280_MEAS_DUR * hum_osr) + BME280_PRES_HUM_MEAS_OFFSET)) / BME280_MEAS_SCALING_FACTOR);
return max_delay;
}
/*!
* @brief This internal API sets the oversampling settings for pressure,
* temperature and humidity in the sensor.
*/
static int8_t set_osr_settings(uint8_t desired_settings,
const struct bme280_settings *settings,
const struct bme280_dev *dev)
{
int8_t rslt = BME280_W_INVALID_OSR_MACRO;
if (desired_settings & BME280_OSR_HUM_SEL)
{
rslt = set_osr_humidity_settings(settings, dev);
}
if (desired_settings & (BME280_OSR_PRESS_SEL | BME280_OSR_TEMP_SEL))
{
rslt = set_osr_press_temp_settings(desired_settings, settings, dev);
}
return rslt;
}
/*!
* @brief This API sets the humidity oversampling settings of the sensor.
*/
static int8_t set_osr_humidity_settings(const struct bme280_settings *settings, const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t ctrl_hum;
uint8_t ctrl_meas;
uint8_t reg_addr = BME280_CTRL_HUM_ADDR;
ctrl_hum = settings->osr_h & BME280_CTRL_HUM_MSK;
/* Write the humidity control value in the register */
rslt = bme280_set_regs(&reg_addr, &ctrl_hum, 1, dev);
/* Humidity related changes will be only effective after a
* write operation to ctrl_meas register
*/
if (rslt == BME280_OK)
{
reg_addr = BME280_CTRL_MEAS_ADDR;
rslt = bme280_get_regs(reg_addr, &ctrl_meas, 1, dev);
if (rslt == BME280_OK)
{
rslt = bme280_set_regs(&reg_addr, &ctrl_meas, 1, dev);
}
}
return rslt;
}
/*!
* @brief This API sets the pressure and/or temperature oversampling settings
* in the sensor according to the settings selected by the user.
*/
static int8_t set_osr_press_temp_settings(uint8_t desired_settings,
const struct bme280_settings *settings,
const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t reg_addr = BME280_CTRL_MEAS_ADDR;
uint8_t reg_data;
rslt = bme280_get_regs(reg_addr, &reg_data, 1, dev);
if (rslt == BME280_OK)
{
if (desired_settings & BME280_OSR_PRESS_SEL)
{
fill_osr_press_settings(&reg_data, settings);
}
if (desired_settings & BME280_OSR_TEMP_SEL)
{
fill_osr_temp_settings(&reg_data, settings);
}
/* Write the oversampling settings in the register */
rslt = bme280_set_regs(&reg_addr, &reg_data, 1, dev);
}
return rslt;
}
/*!
* @brief This internal API sets the filter and/or standby duration settings
* in the sensor according to the settings selected by the user.
*/
static int8_t set_filter_standby_settings(uint8_t desired_settings,
const struct bme280_settings *settings,
const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t reg_addr = BME280_CONFIG_ADDR;
uint8_t reg_data;
rslt = bme280_get_regs(reg_addr, &reg_data, 1, dev);
if (rslt == BME280_OK)
{
if (desired_settings & BME280_FILTER_SEL)
{
fill_filter_settings(&reg_data, settings);
}
if (desired_settings & BME280_STANDBY_SEL)
{
fill_standby_settings(&reg_data, settings);
}
/* Write the oversampling settings in the register */
rslt = bme280_set_regs(&reg_addr, &reg_data, 1, dev);
}
return rslt;
}
/*!
* @brief This internal API fills the filter settings provided by the user
* in the data buffer so as to write in the sensor.
*/
static void fill_filter_settings(uint8_t *reg_data, const struct bme280_settings *settings)
{
*reg_data = BME280_SET_BITS(*reg_data, BME280_FILTER, settings->filter);
}
/*!
* @brief This internal API fills the standby duration settings provided by
* the user in the data buffer so as to write in the sensor.
*/
static void fill_standby_settings(uint8_t *reg_data, const struct bme280_settings *settings)
{
*reg_data = BME280_SET_BITS(*reg_data, BME280_STANDBY, settings->standby_time);
}
/*!
* @brief This internal API fills the pressure oversampling settings provided by
* the user in the data buffer so as to write in the sensor.
*/
static void fill_osr_press_settings(uint8_t *reg_data, const struct bme280_settings *settings)
{
*reg_data = BME280_SET_BITS(*reg_data, BME280_CTRL_PRESS, settings->osr_p);
}
/*!
* @brief This internal API fills the temperature oversampling settings
* provided by the user in the data buffer so as to write in the sensor.
*/
static void fill_osr_temp_settings(uint8_t *reg_data, const struct bme280_settings *settings)
{
*reg_data = BME280_SET_BITS(*reg_data, BME280_CTRL_TEMP, settings->osr_t);
}
/*!
* @brief This internal API parse the oversampling(pressure, temperature
* and humidity), filter and standby duration settings and store in the
* device structure.
*/
static void parse_device_settings(const uint8_t *reg_data, struct bme280_settings *settings)
{
settings->osr_h = BME280_GET_BITS_POS_0(reg_data[0], BME280_CTRL_HUM);
settings->osr_p = BME280_GET_BITS(reg_data[2], BME280_CTRL_PRESS);
settings->osr_t = BME280_GET_BITS(reg_data[2], BME280_CTRL_TEMP);
settings->filter = BME280_GET_BITS(reg_data[3], BME280_FILTER);
settings->standby_time = BME280_GET_BITS(reg_data[3], BME280_STANDBY);
}
/*!
* @brief This internal API writes the power mode in the sensor.
*/
static int8_t write_power_mode(uint8_t sensor_mode, const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t reg_addr = BME280_PWR_CTRL_ADDR;
/* Variable to store the value read from power mode register */
uint8_t sensor_mode_reg_val;
/* Read the power mode register */
rslt = bme280_get_regs(reg_addr, &sensor_mode_reg_val, 1, dev);
/* Set the power mode */
if (rslt == BME280_OK)
{
sensor_mode_reg_val = BME280_SET_BITS_POS_0(sensor_mode_reg_val, BME280_SENSOR_MODE, sensor_mode);
/* Write the power mode in the register */
rslt = bme280_set_regs(&reg_addr, &sensor_mode_reg_val, 1, dev);
}
return rslt;
}
/*!
* @brief This internal API puts the device to sleep mode.
*/
static int8_t put_device_to_sleep(const struct bme280_dev *dev)
{
int8_t rslt;
uint8_t reg_data[4];
struct bme280_settings settings;
rslt = bme280_get_regs(BME280_CTRL_HUM_ADDR, reg_data, 4, dev);
if (rslt == BME280_OK)
{
parse_device_settings(reg_data, &settings);
rslt = bme280_soft_reset(dev);
if (rslt == BME280_OK)
{
rslt = reload_device_settings(&settings, dev);
}
}
return rslt;
}
/*!
* @brief This internal API reloads the already existing device settings in
* the sensor after soft reset.
*/
static int8_t reload_device_settings(const struct bme280_settings *settings, const struct bme280_dev *dev)
{
int8_t rslt;
rslt = set_osr_settings(BME280_ALL_SETTINGS_SEL, settings, dev);
if (rslt == BME280_OK)
{
rslt = set_filter_standby_settings(BME280_ALL_SETTINGS_SEL, settings, dev);
}
return rslt;
}
#ifdef BME280_FLOAT_ENABLE
/*!
* @brief This internal API is used to compensate the raw temperature data and
* return the compensated temperature data in double data type.
*/
static double compensate_temperature(const struct bme280_uncomp_data *uncomp_data, struct bme280_calib_data *calib_data)
{
double var1;
double var2;
double temperature;
double temperature_min = -40;
double temperature_max = 85;
var1 = ((double)uncomp_data->temperature) / 16384.0 - ((double)calib_data->dig_t1) / 1024.0;
var1 = var1 * ((double)calib_data->dig_t2);
var2 = (((double)uncomp_data->temperature) / 131072.0 - ((double)calib_data->dig_t1) / 8192.0);
var2 = (var2 * var2) * ((double)calib_data->dig_t3);
calib_data->t_fine = (int32_t)(var1 + var2);
temperature = (var1 + var2) / 5120.0;
if (temperature < temperature_min)
{
temperature = temperature_min;
}
else if (temperature > temperature_max)
{
temperature = temperature_max;
}
return temperature;
}
/*!
* @brief This internal API is used to compensate the raw pressure data and
* return the compensated pressure data in double data type.
*/
static double compensate_pressure(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data)
{
double var1;
double var2;
double var3;
double pressure;
double pressure_min = 30000.0;
double pressure_max = 110000.0;
var1 = ((double)calib_data->t_fine / 2.0) - 64000.0;
var2 = var1 * var1 * ((double)calib_data->dig_p6) / 32768.0;
var2 = var2 + var1 * ((double)calib_data->dig_p5) * 2.0;
var2 = (var2 / 4.0) + (((double)calib_data->dig_p4) * 65536.0);
var3 = ((double)calib_data->dig_p3) * var1 * var1 / 524288.0;
var1 = (var3 + ((double)calib_data->dig_p2) * var1) / 524288.0;
var1 = (1.0 + var1 / 32768.0) * ((double)calib_data->dig_p1);
/* avoid exception caused by division by zero */
if (var1 > (0.0))
{
pressure = 1048576.0 - (double) uncomp_data->pressure;
pressure = (pressure - (var2 / 4096.0)) * 6250.0 / var1;
var1 = ((double)calib_data->dig_p9) * pressure * pressure / 2147483648.0;
var2 = pressure * ((double)calib_data->dig_p8) / 32768.0;
pressure = pressure + (var1 + var2 + ((double)calib_data->dig_p7)) / 16.0;
if (pressure < pressure_min)
{
pressure = pressure_min;
}
else if (pressure > pressure_max)
{
pressure = pressure_max;
}
}
else /* Invalid case */
{
pressure = pressure_min;
}
return pressure;
}
/*!
* @brief This internal API is used to compensate the raw humidity data and
* return the compensated humidity data in double data type.
*/
static double compensate_humidity(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data)
{
double humidity;
double humidity_min = 0.0;
double humidity_max = 100.0;
double var1;
double var2;
double var3;
double var4;
double var5;
double var6;
var1 = ((double)calib_data->t_fine) - 76800.0;
var2 = (((double)calib_data->dig_h4) * 64.0 + (((double)calib_data->dig_h5) / 16384.0) * var1);
var3 = uncomp_data->humidity - var2;
var4 = ((double)calib_data->dig_h2) / 65536.0;
var5 = (1.0 + (((double)calib_data->dig_h3) / 67108864.0) * var1);
var6 = 1.0 + (((double)calib_data->dig_h6) / 67108864.0) * var1 * var5;
var6 = var3 * var4 * (var5 * var6);
humidity = var6 * (1.0 - ((double)calib_data->dig_h1) * var6 / 524288.0);
if (humidity > humidity_max)
{
humidity = humidity_max;
}
else if (humidity < humidity_min)
{
humidity = humidity_min;
}
return humidity;
}
#else
/*!
* @brief This internal API is used to compensate the raw temperature data and
* return the compensated temperature data in integer data type.
*/
static int32_t compensate_temperature(const struct bme280_uncomp_data *uncomp_data,
struct bme280_calib_data *calib_data)
{
int32_t var1;
int32_t var2;
int32_t temperature;
int32_t temperature_min = -4000;
int32_t temperature_max = 8500;
var1 = (int32_t)((uncomp_data->temperature / 8) - ((int32_t)calib_data->dig_t1 * 2));
var1 = (var1 * ((int32_t)calib_data->dig_t2)) / 2048;
var2 = (int32_t)((uncomp_data->temperature / 16) - ((int32_t)calib_data->dig_t1));
var2 = (((var2 * var2) / 4096) * ((int32_t)calib_data->dig_t3)) / 16384;
calib_data->t_fine = var1 + var2;
temperature = (calib_data->t_fine * 5 + 128) / 256;
if (temperature < temperature_min)
{
temperature = temperature_min;
}
else if (temperature > temperature_max)
{
temperature = temperature_max;
}
return temperature;
}
#ifdef BME280_64BIT_ENABLE
/*!
* @brief This internal API is used to compensate the raw pressure data and
* return the compensated pressure data in integer data type with higher
* accuracy.
*/
static uint32_t compensate_pressure(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data)
{
int64_t var1;
int64_t var2;
int64_t var3;
int64_t var4;
uint32_t pressure;
uint32_t pressure_min = 3000000;
uint32_t pressure_max = 11000000;
var1 = ((int64_t)calib_data->t_fine) - 128000;
var2 = var1 * var1 * (int64_t)calib_data->dig_p6;
var2 = var2 + ((var1 * (int64_t)calib_data->dig_p5) * 131072);
var2 = var2 + (((int64_t)calib_data->dig_p4) * 34359738368);
var1 = ((var1 * var1 * (int64_t)calib_data->dig_p3) / 256) + ((var1 * ((int64_t)calib_data->dig_p2) * 4096));
var3 = ((int64_t)1) * 140737488355328;
var1 = (var3 + var1) * ((int64_t)calib_data->dig_p1) / 8589934592;
/* To avoid divide by zero exception */
if (var1 != 0)
{
var4 = 1048576 - uncomp_data->pressure;
var4 = (((var4 * INT64_C(2147483648)) - var2) * 3125) / var1;
var1 = (((int64_t)calib_data->dig_p9) * (var4 / 8192) * (var4 / 8192)) / 33554432;
var2 = (((int64_t)calib_data->dig_p8) * var4) / 524288;
var4 = ((var4 + var1 + var2) / 256) + (((int64_t)calib_data->dig_p7) * 16);
pressure = (uint32_t)(((var4 / 2) * 100) / 128);
if (pressure < pressure_min)
{
pressure = pressure_min;
}
else if (pressure > pressure_max)
{
pressure = pressure_max;
}
}
else
{
pressure = pressure_min;
}
return pressure;
}
#else
/*!
* @brief This internal API is used to compensate the raw pressure data and
* return the compensated pressure data in integer data type.
*/
static uint32_t compensate_pressure(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data)
{
int32_t var1;
int32_t var2;
int32_t var3;
int32_t var4;
uint32_t var5;
uint32_t pressure;
uint32_t pressure_min = 30000;
uint32_t pressure_max = 110000;
var1 = (((int32_t)calib_data->t_fine) / 2) - (int32_t)64000;
var2 = (((var1 / 4) * (var1 / 4)) / 2048) * ((int32_t)calib_data->dig_p6);
var2 = var2 + ((var1 * ((int32_t)calib_data->dig_p5)) * 2);
var2 = (var2 / 4) + (((int32_t)calib_data->dig_p4) * 65536);
var3 = (calib_data->dig_p3 * (((var1 / 4) * (var1 / 4)) / 8192)) / 8;
var4 = (((int32_t)calib_data->dig_p2) * var1) / 2;
var1 = (var3 + var4) / 262144;
var1 = (((32768 + var1)) * ((int32_t)calib_data->dig_p1)) / 32768;
/* avoid exception caused by division by zero */
if (var1)
{
var5 = (uint32_t)((uint32_t)1048576) - uncomp_data->pressure;
pressure = ((uint32_t)(var5 - (uint32_t)(var2 / 4096))) * 3125;
if (pressure < 0x80000000)
{
pressure = (pressure << 1) / ((uint32_t)var1);
}
else
{
pressure = (pressure / (uint32_t)var1) * 2;
}
var1 = (((int32_t)calib_data->dig_p9) * ((int32_t)(((pressure / 8) * (pressure / 8)) / 8192))) / 4096;
var2 = (((int32_t)(pressure / 4)) * ((int32_t)calib_data->dig_p8)) / 8192;
pressure = (uint32_t)((int32_t)pressure + ((var1 + var2 + calib_data->dig_p7) / 16));
if (pressure < pressure_min)
{
pressure = pressure_min;
}
else if (pressure > pressure_max)
{
pressure = pressure_max;
}
}
else
{
pressure = pressure_min;
}
return pressure;
}
#endif
/*!
* @brief This internal API is used to compensate the raw humidity data and
* return the compensated humidity data in integer data type.
*/
static uint32_t compensate_humidity(const struct bme280_uncomp_data *uncomp_data,
const struct bme280_calib_data *calib_data)
{
int32_t var1;
int32_t var2;
int32_t var3;
int32_t var4;
int32_t var5;
uint32_t humidity;
uint32_t humidity_max = 102400;
var1 = calib_data->t_fine - ((int32_t)76800);
var2 = (int32_t)(uncomp_data->humidity * 16384);
var3 = (int32_t)(((int32_t)calib_data->dig_h4) * 1048576);
var4 = ((int32_t)calib_data->dig_h5) * var1;
var5 = (((var2 - var3) - var4) + (int32_t)16384) / 32768;
var2 = (var1 * ((int32_t)calib_data->dig_h6)) / 1024;
var3 = (var1 * ((int32_t)calib_data->dig_h3)) / 2048;
var4 = ((var2 * (var3 + (int32_t)32768)) / 1024) + (int32_t)2097152;
var2 = ((var4 * ((int32_t)calib_data->dig_h2)) + 8192) / 16384;
var3 = var5 * var2;
var4 = ((var3 / 32768) * (var3 / 32768)) / 128;
var5 = var3 - ((var4 * ((int32_t)calib_data->dig_h1)) / 16);
var5 = (var5 < 0 ? 0 : var5);
var5 = (var5 > 419430400 ? 419430400 : var5);
humidity = (uint32_t)(var5 / 4096);
if (humidity > humidity_max)
{
humidity = humidity_max;
}
return humidity;
}
#endif
/*!
* @brief This internal API reads the calibration data from the sensor, parse
* it and store in the device structure.
*/
static int8_t get_calib_data(struct bme280_dev *dev)
{
int8_t rslt;
uint8_t reg_addr = BME280_TEMP_PRESS_CALIB_DATA_ADDR;
/* Array to store calibration data */
uint8_t calib_data[BME280_TEMP_PRESS_CALIB_DATA_LEN] = { 0 };
/* Read the calibration data from the sensor */
rslt = bme280_get_regs(reg_addr, calib_data, BME280_TEMP_PRESS_CALIB_DATA_LEN, dev);
if (rslt == BME280_OK)
{
/* Parse temperature and pressure calibration data and store
* it in device structure
*/
parse_temp_press_calib_data(calib_data, dev);
reg_addr = BME280_HUMIDITY_CALIB_DATA_ADDR;
/* Read the humidity calibration data from the sensor */
rslt = bme280_get_regs(reg_addr, calib_data, BME280_HUMIDITY_CALIB_DATA_LEN, dev);
if (rslt == BME280_OK)
{
/* Parse humidity calibration data and store it in
* device structure
*/
parse_humidity_calib_data(calib_data, dev);
}
}
return rslt;
}
/*!
* @brief This internal API interleaves the register address between the
* register data buffer for burst write operation.
*/
static void interleave_reg_addr(const uint8_t *reg_addr, uint8_t *temp_buff, const uint8_t *reg_data, uint8_t len)
{
uint8_t index;
for (index = 1; index < len; index++)
{
temp_buff[(index * 2) - 1] = reg_addr[index];
temp_buff[index * 2] = reg_data[index];
}
}
/*!
* @brief This internal API is used to parse the temperature and
* pressure calibration data and store it in device structure.
*/
static void parse_temp_press_calib_data(const uint8_t *reg_data, struct bme280_dev *dev)
{
struct bme280_calib_data *calib_data = &dev->calib_data;
calib_data->dig_t1 = BME280_CONCAT_BYTES(reg_data[1], reg_data[0]);
calib_data->dig_t2 = (int16_t)BME280_CONCAT_BYTES(reg_data[3], reg_data[2]);
calib_data->dig_t3 = (int16_t)BME280_CONCAT_BYTES(reg_data[5], reg_data[4]);
calib_data->dig_p1 = BME280_CONCAT_BYTES(reg_data[7], reg_data[6]);
calib_data->dig_p2 = (int16_t)BME280_CONCAT_BYTES(reg_data[9], reg_data[8]);
calib_data->dig_p3 = (int16_t)BME280_CONCAT_BYTES(reg_data[11], reg_data[10]);
calib_data->dig_p4 = (int16_t)BME280_CONCAT_BYTES(reg_data[13], reg_data[12]);
calib_data->dig_p5 = (int16_t)BME280_CONCAT_BYTES(reg_data[15], reg_data[14]);
calib_data->dig_p6 = (int16_t)BME280_CONCAT_BYTES(reg_data[17], reg_data[16]);
calib_data->dig_p7 = (int16_t)BME280_CONCAT_BYTES(reg_data[19], reg_data[18]);
calib_data->dig_p8 = (int16_t)BME280_CONCAT_BYTES(reg_data[21], reg_data[20]);
calib_data->dig_p9 = (int16_t)BME280_CONCAT_BYTES(reg_data[23], reg_data[22]);
calib_data->dig_h1 = reg_data[25];
}
/*!
* @brief This internal API is used to parse the humidity calibration data
* and store it in device structure.
*/
static void parse_humidity_calib_data(const uint8_t *reg_data, struct bme280_dev *dev)
{
struct bme280_calib_data *calib_data = &dev->calib_data;
int16_t dig_h4_lsb;
int16_t dig_h4_msb;
int16_t dig_h5_lsb;
int16_t dig_h5_msb;
calib_data->dig_h2 = (int16_t)BME280_CONCAT_BYTES(reg_data[1], reg_data[0]);
calib_data->dig_h3 = reg_data[2];
dig_h4_msb = (int16_t)(int8_t)reg_data[3] * 16;
dig_h4_lsb = (int16_t)(reg_data[4] & 0x0F);
calib_data->dig_h4 = dig_h4_msb | dig_h4_lsb;
dig_h5_msb = (int16_t)(int8_t)reg_data[5] * 16;
dig_h5_lsb = (int16_t)(reg_data[4] >> 4);
calib_data->dig_h5 = dig_h5_msb | dig_h5_lsb;
calib_data->dig_h6 = (int8_t)reg_data[6];
}
/*!
* @brief This internal API is used to identify the settings which the user
* wants to modify in the sensor.
*/
static uint8_t are_settings_changed(uint8_t sub_settings, uint8_t desired_settings)
{
uint8_t settings_changed = FALSE;
if (sub_settings & desired_settings)
{
/* User wants to modify this particular settings */
settings_changed = TRUE;
}
else
{
/* User don't want to modify this particular settings */
settings_changed = FALSE;
}
return settings_changed;
}
/*!
* @brief This internal API is used to validate the device structure pointer for
* null conditions.
*/
static int8_t null_ptr_check(const struct bme280_dev *dev)
{
int8_t rslt;
if ((dev == NULL) || (dev->read == NULL) || (dev->write == NULL) || (dev->delay_ms == NULL))
{
/* Device structure pointer is not valid */
rslt = BME280_E_NULL_PTR;
}
else
{
/* Device structure is fine */
rslt = BME280_OK;
}
return rslt;
}